Contact an Account ManagerSeparation of pesticides according to DIN EN ISO 11369 (F12)
Summary
Type: HPLC
Application #: 126460
Catalog#: 760052.40 - MN HPLC Column, Nucleodur C18 ec, 100A, 3um, 4x250mm
Detection:
Conditions: Eluent A: acetonitrile
Eluent B: acetonitrile – 1 mmol/L ammonium acetate (12.5:87.5, v/v)
Gradient: 100 % B in 60 min to 55 % B, in 9 min to 10 % B
Flow rate: 0.50 mL/min
Temperature: 40 °C
Injection volume: 40 µL
Authors: Coelhan, M.
Source: Forschungszentrum Weihenstephan, Germany 2013